 Tischversion
 Wall Version
19'' Rack version available
Applications:
Applications are in the process analyzers for GC FID, THC, NOX, BTX AND SOX analyzers.
Gas Characteristics:
Operating pressure: 2-8 barg
Total Hydrocarbons residual < 0,1 ppm
HC removal efficiency:
> 99% methane (CH4)
> 99,5% aromatics (BTX) and hydrocarbons starting from C3Hn
> 99,9% carbon monoxide (CO), hydrogen (H2) ethylene and propylene
Air Supply Characteristics:
Dew point: < +3 °C
Particles: < 0,5 μm
Oil vapor: < 0,01 mg/m3
|
Claind's zero air purifiers produce ultra high pure zero air from a compressed air supply. Different models, designed for every kind of laboratory, are realized in rack, desk and wall version with output flow rate from 1 to 20 l/min.
Total hydrocarbons (THC) are reduced under 0.1 ppm allowing extremely low, flat and stable baseline of GC-FID analyzers, enhancing sensitivity and improving report analysis.
All is needed is a mains power supply, and a compressed air supply to produce ultra high purity zero air 24 hours per day. An internal temperature controller will maintain constantly the catalytic reactor to its optimum operating temperature.
The heart of Claind’s zero air purifiers is a platinum-palladium catalytic reactor, designed and manufactured by Claind for its exclusive use.
Working Principle:
The elimination of hydrocarbon, till less than 0.1 ppm, takes place through the strong oxidant power of the platinum and palladium catalytic reactor, which is maintained at the temperature of 385° C.
Optional filters, put in series mode with the reactor, allow the adsorption of the sulphurous (SOX) and nitrogenous (NOX) products and H2S and O3.
The hydrocarbon elimination is with efficiency >99 % for CH4 e > 99.5 % for BX, TX, C3H4, ETYL, PROPYL, and so on.
Technical Specifications and Dimensions |
Modell* | Flow rate Nl/min | Power VA
| Height cm | Width cm | Depth cm | Weight kg |
AZ2001 W | 1 | 120 | 32 | 15 | 28 | 7 |
AZ2005 W | 5 | 120 | 32 | 15 | 28 | 7 |
AZ2005 R | 5 | 120 | 13 (3HE) | 48 (19'') | 27 | 8 |
AZ2010 R | 10 | 170 | 18 (4HE) | 48 (19”) | 27 | 10 |
AZ2020 R | 20 | 170 | 18 (4HE) | 48 (19”) | 27 | 10 |
AZ2005 B | 5 | 120 | 40 | 42 | 32 | 13 |
AZ2010 B | 10 | 170 | 40 | 42 | 32 | 13 |
AZ2020 B | 20 | 170 | 40 | 42 | 32 | 13 |
* R=RACK, B=DESK, W=WALL
Electrical supply: 230 Vac 50 Hz, 110 Vac 60Hz
Operating temperature: between 5°C and 40°C
Consumables: none
|