Pure Gas Products

Multiple Gas Generators

English
Gas Generators
   

CLAIND Laboratory Gas Generators - Multiple Gas Generator ANG2000 - Nitrogen and Air


Applications
  • gas chromatography to feed ECD, carrier gas, MAKE UP for detectors
  • Thermoanalysis
  • Sample preparation
The ANG generator can supply contemporarily and constantly up to 3 Nl/min of zero air and up to 800 Nml/min of nitrogen at 6 barg pressure and purity better than 99,9995%.
The nitrogen side can be equipped with a catalytic reactor to reduce hydrocarbons to values lower than 0,1 ppm (HC feature).
Independence is one of the main characteristics of these generators. They only require a supply of mains electricity for automatic, unattended operation. Pressing a button is the only operation to start up the generator, which operates automatically 24 hours a day, 365 days a year.
Maintenance is limited to changing a few filter elements taking less than 30 minutes per year. The system regenerates itself automatically giving you a continuous, uninterrupted supply of gas.

Working Principle:
Both air and nitrogen are produced according to the PSA principle (Pressure Swing Adsorption).
Ambient air is purified from oxygen, moisture, CO and other ‘contaminants’ by passing through a Carbon Molecular Sieve filter (CMS).
By using different composition of CMS and by setting different operating parameters, purified air is obtained from one side and high pure nitrogen from the other side.
HC removal is done with a catalytic reactor (only nitrogen side).

Main Features
  • Built in compressor
  • Fast start™ feature to allow the maximum purity just few minutes after switching on
  • Stand by mode allows to produce the only air required
  • Compact instrument, just 40 cm wide
  • Catalytic reactor for hydrocarbons removal
gas characteristics: Outlet pressure: 6 barg
Purity
ANG 2381ANG 2381HC
Nitrogen purity99,9995%99,9995%
Moisture content AIR side< -60 °C dew point< -60 °C dew point
Total hydrocarbons residual*...< 0,1 ppm
* tested in Claind laboratories in standard conditions @ 25°C

HC feature (hydrocarbons removal):
HC removal efficiency:
> 99% methane (CH4)
> 99,5% aromatics (BTX) and hydrocarbons starting from C3Hn
> 99,9% carbon monoxide (CO), hydrogen (H2) ethylene and propylene
Technical Specifications
ANG 2381ANG 2381HC
N2 Flow rate800 Nml/min800 Nml/min
Air flow rate3000 Nml/min3000 Nml/min
Power600 W700 W
Electrical supply230/110 Vac – 50/60 Hz230/110 Vac – 50/60 Hz

Gas Characteristics:
Outlet pressure: 6 barg
Noise level: < 55 dB
Operating temperature: between 5°C and 40°C
Consumables: filters and silencers
Dimensions
ANG 2381ANG 2381HC
Height83 cm83 cm
Width40 cm40 cm
Depth58 cm58 cm
Weight73 kg88 kg

© 2025 Rainer Lammertz
DisclaimerImprintSitemap