Purge Gas Generator Systems

The Purge Gas Generator Systems are designed to continuously remove water vapor and CO2 gas from compressed air and inert gases. The units use 13X molecular sieve media with the pressure-swing adsorption process to reduce levels of CO2 to below 1 ppm and reduce water vapor content to less than -73°C dew point. The dual tower design provides continuous uninterrupted operation. CO2-Series requires a compressed air source for operation and the air should be compatible with instrument quality air. These gas generator systems are designed to operate 24 hours per day, 7 days a week. Nitrogen cylinders and dewars can be eliminated with the CO2-PG series gas generator systems.
PURGE GAS GENERATORS SPECIFICATIONS
Model No. TOC-PG3 CO2-PG14 CO2-PG28 CO2-PG80
Flow Capacities at 6,9 Bar (lit/min)
Inlet Flow 13,2 25,1 40 95
Purge Flow 5,5 10,3 17 35
Outlet Flow 7,7 14,8 23 60
*Flow Capacities at 8,6 Bar (lit/min)
Inlet Flow - - 48 114
Purge Flow - - 16 34
Outlet Flow - - 32 80
Connections (NPT)
Inlet 1/8" 1/4" 1/4" 1/4"
Outlet 1/8" 1/4" 1/4" 1/4"
Dimensions (cm)
Height 27 32 50 70
Width 32 41 44 44
Depth 11 13 16 16
Weight (kg) 2,9 4,9 9,1 11,3
Contact factory for additional flow capacities

FEATURES
  • Reduces CO2 to less than 1ppm
  • Reduces water vapor to less than 1 ppm
  • Compact, lightweight design saves space and is easy to install
  • Inlet filter 0.01 micron coalescing with automatic drain included
  • Outlet filter/regulator with gauge included
  • Operating pressure 3,4 to 8,3 Bar (14 lit/min)
  • Operating pressure 3,4 to 10 Bar (28 & 80 lit/min)
  • Continuous self-regenerative operation
  • Sound insulated enclosure. 115 Volts AC, 6 watts
  • Power cord with plug included

TYPICAL APPLICATIONS
  • FTIR Spectrometers
  • TOC Analyzers
  • Continuous Emissions Monitors
  • Gas Chromatographs
  • Laboratory Analyzers

DOWNLOADS