Tracer 2™

Tracer 2™

The Tracer2™ addresses the need for accurate, reliable process controls, and moisture sampling at levels at or below single-digit parts-per-billion. It is ideally suited for ultra-high-purity moisture detection in semiconductor manufacturing processes, inert and specialty gases, component and system testing, and laboratory research applications.
Based on Faraday's Law of Electrolysis, the TRACER 2 sensor absorbs and electrolyzes moisture down to fractional ppb's. How? One hundred percent of the sample moisture is absorbed by aphosphorus pentoxide (P2O5) film that covers two spirally-wound electrodes embed in a hollow glass tube

Tracer 2™ Specifications:
Operating range:Inerts: 0 - 2000 ppb
Oxygen: 0 - 1200 ppb
Hydrogen: 0 - 500 ppb
Cell - 700 ppt LDL
Sensor:MEECO Counter flow Cell
(U.S. Patent No.s 5225065 and 5198094)
Accuracy:Inert Mode: 4% of reading or ± 4 ppb, whichever is greater.
H2/O2 Mode: When the value is < 50 ppb, the accuracy is 6% of reading or ±6 ppb, whichever is greater.
When the value is >50 ppb, the accuracy is 2% of reading or ±20 ppb, whichever is greater.
Alarms:Two configurable alarms. Can be set anywhere within the operating range.
Alternately, use it as a responsiveness check or gas trend indicator. One system alarm for cell responsiveness and flow.
Flow Control:Provided by NIST traceable mass flow controllers.
Required Flow: 1200 SCCM minimum
Low Detection Limit:1 ppb at 95% Confidence Limit
Mounting:Standard 7" X 19" Rack Mount for up to two TRACER 2 units
Environmental Conditions:+10° to +40° Celsius
+50° to +104° Fahrenheit
20% to 80% RH, non-condensing, non-corrosive atmosphere
Optimal performance is achieved when temperature fluctuation is maintained below +/-2°C.
Sample Gas:Air, Argon, Carbon Dioxide, Carbon Tetrafluoride*, Helium, Hydrogen, Methane*, Neon, Nitrogen, Krypton, Sulfur Hexafluoride, Oxygen, Xenon, non-corrosive/non- toxic gases.
(* these gases may require special handling. Please consult factory before use.
Sample Inlet Pressure:30 - 150 psig (2 -10.2 Bar)**
**For lower pressure, consult factory
Sensitivity:0.5 ppb or below
Signal Output:One 0-5V DC and 4-20m A analog output: user defied and scaled.
RS-232 serial communication port, two level and one system alarm
User Interface:2-line X 20 character Vacuum
Fluorescent Display
5-key Membrane Keypad
Speed Response:Response to 5 ppb intrusion is 50% in 15 minutes.
Configuration:TRACER 2 DUO: Two TRACER 2s in one 7" X 19" rack mount box.
© 2022 Rainer Lammertz