Peak Performer 1 FID - Specifications

Peak Performer1 FID


Accessories and other features:
  • trend outputs
  • analysis averaging
  • auto start/stop capability
  • EZ Chrom compatible
  • On column syringe injector adapter
  • Dual sample stream option
Accuracy:
  • + / - detection limit or 10% of reading, whichever is higher
Range:
  • 10000 : 1
    Examples: 1 ppb – 10 ppm w/5 cc sample loop
    5 ppb – 50 ppm w/1 cc sample loop
Operation:
  • Run time: 400 seconds < (depending on application)
  • Operating Temperature: 55 - 85 °F (13 - 30 °C )
  • Gas Supply Requirements:
    • Carrier: Purified nitrogen
      Supply pressure 70 – 110 psig with 5% stability
    • FID H2: 99.999% pure or purified: < 1000 ppb impurities
      Supply pressure 25 – 45 psig with 5% stability
    • FID Air: 99.999% pure or purified: < 1000 ppb impurities
      Supply pressure 5 – 45 psig with 5% stability
  • Data Collection / Communication (Standard):
    • Backlit 3” x 5.25” touch screen
    • Visual chromatogram and numerical results
    • On-board integration with re-run capability
    • Chromatograms collected and archived via standard 9 – pin com-port in .csv format (data is excel compatible)
    • Printer port
Dimensions / Electrical:
  • 27” L x 17” W x 7” H
  • 25 lbs.
  • 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz
  • 1.5 amp maximum
© 2022 Rainer Lammertz